Paediatric pulse oximeter with gleeful motifs for the measurement of the oxygen saturation and the pulse in paediatric patients.
For home and professional use.
*For children aged between 3 and 14 or with finger diameter between 8 and 12 mm.
FEATURES:
• OLED screen.
• SpO2 measurement, HR and PI.
• Plethysmogram.
• Automatic start and turn off.
• Low-battery indicator.
• Display of 4 directions,just by pressing the button, for a perfect reading.
TECHNICAL SPECIFICATIONS:
• SpO2 measurement.
Transducer: dual wavelength LED sensor with wavelength:
Red light: 663 nm, infrared light: 890 nm.
Maximum power of average optical output: ≤ 2 mW
viewing range of SpO2: 35% ~ 100%
Accuracy in SpO2 measurement: ≤ 3% for a range of a SpO2 from 70% to 100%
Heart rate measurement.
Viewing range of HR: 30 bpm ~ 240 bpm
Accuracy in HR measuring: ± 2 bpm or ± 2% (the biggest one)
Viewing range of the perfusion index (PI) 0% ~ 20%
Preset excessive limits.
Inferior limit of SpO2: 90%
Heart rate: superior limit: 120 bpm
Inferior limit: 50 bpm
• Adjustments of the superior limit:
o SpO2:
Inferior limit by default: 90
Heart rate:
Superior limit: 120 bpm
Inferior limit: 50 bpm
Function of visual and audio alert.
During the measurement, if the SpO2 value or the HR value surpass the preset limit, the device will automatically warn you with a beep, and the exceeding value will flicker on the display.
• Power source requirements:
2 alkaline batteries LR03 (AAA)
Voltage: 3.0V DC
Power: ≤ 40 mA
Environment conditions:
Temperature: 5°C ~ 40°C
Humidity: 30% ~ 80%
Atmospheric pressure: 70 kPa ~ 106 kPa
Low perfusion performance:
The accuracy in the SpO2 and HR measuring continues keeping the formerly indicated precision when the amplitude modulation is 0.6%.
• Ambient light interference:
The difference between the SpO2 value measured in interior places with natural light and the value of a dark room is smaller than ± 1%.
• Sizes: 66 mm (L) × 36 mm (A) × 33 mm (A) Weight Net: approx. 60 g
• Classification:
Type of protection against electric shocks: Internally powered equipment.
Degree of protection against electric shocks: Component of the BF type.
Degree of protection against harmful solid strange objects and the entry of liquids:
The equipment has the protection level IP22 against harmful solid strange objects and the liquids entry.
Electromagnetic compatibility: Group I, Class B
Over
70 years
now we are active in the field of health, with high quality products and services, from the largest domestic and international health product houses!